C Mixture cleaning C v1.00 C The flow valve with V12 needs to be adjusted properly C Suggested: one turn more closed than normal C Also the script requires the mixture to be fully C in the dumps at the start C var1 is initial p5 C var2 is cycle initial p5 C var3 is the number of cycles completed C var4 is cycle initial p2 C var5 is cycle initial time START STATE 1 SET_STATE 100 STATE 100 IF (turbo1==1)||(scroll2==1)||(v15==1)||(v17==1)||(v18==1)||(v1==1) ERROR Stop turbo and disconnect from aux C Stop turbo and disconnect from aux ELSE SET_STATE 101 ENDIF STATE 101 OFF_ALL SET_STATE 102 STATE 102 ON scroll1 SET_STATE 103 WAIT 5 STATE 103 Open Valves C Adjust Valves ON v2 ON v3 ON v4 SET_STATE 104 WAIT 1 STATE 104 Check Pressures IF (p2>0.2)||(p3>2) ERROR Remove mixture from system first C Remove mixture from system first ELSE SET_STATE 200 ENDIF C Cleaning sequence (x5 see state 265) STATE 200 Initialize cleaning SET var1=p5 SET var3=0 SET_STATE 201 STATE 201 Cycle Start OFF v7 SET_STATE 202 WAIT 1 STATE 202 ON v9 C Wait for mixture to cooldown in trap SET_STATE 203 WAIT 30 STATE 203 ON v7 SET_STATE 205 WAIT 60 STATE 205 SET var2=p5 SET var4=p2 SET var5=time ON v12 SET_STATE 210 WAIT 1 STATE 210 Rate checks manual valve closed IF p5<400 OFF v12 ERROR Manual dump valve is probably closed ELSE SET_STATE 211 ENDIF STATE 211 Rate checks trap plugged IF p4>(p2+50) OFF v12 ERROR Traps are probably plugged. Clean them! ELSE SET_STATE 212 ENDIF STATE 212 Rate checks p2 increase pressure C test for first minute IF (p2>p5)&&((time-var5)<60) OFF v12 ERROR Flow valve proabably too open. p2 increases too rapidly. ELSE SET_STATE 213 ENDIF STATE 213 Rate checks p2 increase rate C Limit of 80 mbar/min after first 30s IF ((time-var5)>30)&&(((p2-var4)/(time-var5)*60)>80) OFF v12 ERROR Flow valve proabably too open. p2 increases too rapidly. ELSE SET_STATE 214 ENDIF STATE 214 Rate checks p5 decrease rate C Limit of 40 mbar/min after first 30s IF ((time-var5)>30)&&(((var2-p5)/(time-var5)*60)>40) OFF v12 ERROR Flow valve proabably too open or Manual dump valve closed. p5 goes down too fast. ELSE SET_STATE 240 ENDIF STATE 240 Check End of cycle IF p2>800 SET_STATE 250 ELSE C Continue Checking SET_STATE 210 WAIT 5 ENDIF STATE 250 OFF v12 OFF v9 SET_STATE 255 WAIT 1 STATE 255 ON v13 SET_STATE 260 WAIT 10 STATE 260 Pump mixture back into dumps ON v10 SET_STATE 270 STATE 270 Check pumping IF p2<0.4 OFF v10 OFF v13 SET var3=(var3+1) SET_STATE 275 WAIT 15 ELSE SET_STATE 270 WAIT 15 ENDIF STATE 275 Iteration Check IF var3<5 SET_STATE 201 ELSE SET_STATE 300 ENDIF STATE 300 ON v10 ON v13 ON v1 SET_STATE 310 WAIT 30 STATE 310 ON turbo1 OFF v2 SET_STATE 500 STATE 500 STOP END